Journal of equine veterinary science2019; 79; 94-99; doi: 10.1016/j.jevs.2019.05.026

Evaluation of Two Handheld Point-of-Care Blood Gas Analyzers in Healthy Donkeys.

Abstract: There has been increasing interest in blood gas analysis in donkeys. "Point-of-care (POC) testing" is a diagnostic testing performed on or immediately next to the patient. This study assesses the agreement between two POC blood gas analyzers in donkeys. Arterial and venous blood samples were collected from 17 donkeys and analyzed using a fully automated blood gas analyzer (ABL 700 Series Radiometer, Denmark) (RAD) and two POC blood gas analyzers (i-STAT System; VetStat, Idexx). The parameters revealed by all three devices were submitted to a canonical discriminant analysis, to evaluate which parameters differentiated the POC analyzers from the RAD. On the basis of the discriminant analysis, we evaluated the best POC for each parameter registered, in comparison with RAD. The results also changed depending on the type of blood (venous or arterial blood). The agreement between i-STAT and RAD was good for venous samples, but was poor for arterial samples. A poor agreement was found between VetStat and RAD for both venous and arterial samples. The increment of the number of subjects might lead to a better understanding of the potential role of the POCs in clinical setting. Finally, increasing the study population is recommended to set reference values.
Publication Date: 2019-06-08 PubMed ID: 31405509DOI: 10.1016/j.jevs.2019.05.026Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Journal Article
  • Research Support
  • Non-U.S. Gov't

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

This research sought to measure the accuracy of two handheld point-of-care (POC) blood gas analyzers in donkeys. Through collecting and analyzing arterial and venous blood sample from 17 donkeys, the researchers found varied results between the devices and recommended expanding the number of subjects to better understand their potential role and set reference values.

Research Methodologies

  • The research involved 17 healthy donkeys, from which both arterial and venous blood samples were collected.
  • These samples were analyzed using a fully automated blood gas analyzer, referred to as ‘RAD’, alongside two handheld POC blood gas analyzers named i-STAT System and VetStat respectively.
  • As a means of comparison and to measure accuracy, the results from the POC analyzers were assessed against the RAD.

Canonical Discriminant Analysis

  • All parameters revealed by the three devices were put through a canonical discriminant analysis.
  • This process is used to determine which parameters differentiate the POC analyzers from the RAD.
  • Through this process, the most accurate POC for each parameter registered was identified in comparison with the RAD.

Results Analysis

  • It was observed that the results differed depending on whether venous or arterial blood was analyzed.
  • The i-STAT and RAD’s measurements showed good agreement when testing venous samples, but not arterial ones.
  • A poor agreement between VetStat and RAD was noted for both blood types – venous and arterial.

Recommendations

  • The researchers recommend increasing the number of subjects in future studies. This could provide a broader understanding of the potential role of POC blood gas analyzers in clinical practice.
  • Such an expansion of the study population could also help in setting reference values, providing general guidelines for what to expect from these devices in the real-world applications.

Cite This Article

APA
Bonelli F, Laus F, Briganti A, Evangelista F, Bazzano M, Conte G, Sgorbini M. (2019). Evaluation of Two Handheld Point-of-Care Blood Gas Analyzers in Healthy Donkeys. J Equine Vet Sci, 79, 94-99. https://doi.org/10.1016/j.jevs.2019.05.026

Publication

ISSN: 0737-0806
NlmUniqueID: 8216840
Country: United States
Language: English
Volume: 79
Pages: 94-99
PII: S0737-0806(19)30387-9

Researcher Affiliations

Bonelli, Francesca
  • Department of Veterinary Sciences, University of Pisa, San Piero a Grado, Pisa, Italy. Electronic address: francesca.bonelli@unipi.it.
Laus, Fulvio
  • School of Biosciences and Veterinary Medicine, University of Camerino, Macerata, Italy.
Briganti, Angela
  • Department of Veterinary Sciences, University of Pisa, San Piero a Grado, Pisa, Italy.
Evangelista, Flavia
  • Department of Veterinary Sciences, University of Pisa, San Piero a Grado, Pisa, Italy.
Bazzano, Marilena
  • School of Biosciences and Veterinary Medicine, University of Camerino, Macerata, Italy.
Conte, Giuseppe
  • Department of Agriculture, Food and Environment, University of Pisa, Pisa, Italy.
Sgorbini, Micaela
  • Department of Veterinary Sciences, University of Pisa, San Piero a Grado, Pisa, Italy.

MeSH Terms

  • Animals
  • Blood Gas Analysis / veterinary
  • Equidae
  • Point-of-Care Systems
  • Point-of-Care Testing
  • Reference Values

Citations

This article has been cited 2 times.
  1. Barros IO, Sousa RS, Tavares MD, Rêgo RO, Firmino PR, Souza FJA, Abrantes MR, Minervino AHH, Araújo CASC, Ortolani EL, Barrêto Júnior RA. Assessment of Donkey (Equus asinus africanus) Whole Blood Stored in CPDA-1 and CPD/SAG-M Blood Bags. Biology (Basel) 2021 Feb 8;10(2).
    doi: 10.3390/biology10020133pubmed: 33567685google scholar: lookup
  2. Mendoza FJ, Toribio RE. An Overview of Donkey Neonatology. Animals (Basel) 2025 Jul 6;15(13).
    doi: 10.3390/ani15131986pubmed: 40646885google scholar: lookup