Validation and comparison of two methods of measuring lactate in equine plasma.
Abstract: Some methods of lactate (LA) measurement have not been validated appropriately for use in horses. Objective: To validate 2 LA analysers (YSI 2300 Stat Plus and TDx Lactic Acid Assay) for use with equine plasma and to compare plasma [LA] determined by the 2 methods. Methods: Both instruments were evaluated for linearity, parallelism, recovery and precision using serial dilutions of standard LA solutions and equine plasma and then comparing results with linear regression or paired t tests. Plasma [LA] results were compared in 275 blood samples collected from horses exercising at various intensities using Bland-Altman analysis. Level of significance was P < 0.05. Results: YSI exhibited good linearity for both LA standards and equine plasma (P < 0.05) at 0-30 mmol/l. TDx had good linearity at 0-12 mmol/l (P 12 mmol/l and with equine plasma, linearity was decreased. YSI exhibited good parallelism between LA standards and equine plasma LA measurements throughout the 0-30 mmol/l range (P > 0.05). Parallelism was poor with TDx (P < 0.05). Mean ± s.d. % recovery was 101.7 ± 3.4% for YSI (acceptable) and 110.6 ± 8.4% for TDx (unacceptable). Within-run and mean between-run coefficients of variation (CV) for plasma samples tested from 3.3-29.5 mmol/l were 0.4-3.0% for YSI. CVs for samples tested from 2.8-8.0 mmol/l were 17.4-24.1% for TDx. In 275 plasma samples, [LA] ranged from 0.1-42.7 mmol/l and 0.3-50.6 mmol/l for the YSI and TDx methods, respectively. The difference in plasma [LA] determined by the 2 methods was -1.0 ± 3.2 mmol/l, documenting that the TDx overestimated the YSI results by a mean value of 1 mmol/l. Conclusions: It was concluded that the YSI method was a reliable method for measuring equine plasma [LA] from 0-30 mmol/l. The TDx method was found not to be suitable for use with equine plasma due to greater variability in measurements (high CV).
© 2010 EVJ Ltd.
Publication Date: 2011-05-28 PubMed ID: 21058999DOI: 10.1111/j.2042-3306.2010.00219.xGoogle Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
- Comparative Study
- Journal Article
Summary
This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.
The researchers compared two methods of measuring lactate levels in horse plasma and concluded that the YSI method is more reliable while the TDx method showed greater variability and was less suited for this use.
Methods used in the study
- The scientists aimed to validate two lactate analysers (YSI 2300 Stat Plus and TDx Lactic Acid Assay) to assess their suitability for use with equine plasma.
- The measuring instruments were tested for linearity, parallelism, recovery and precision. These evaluations were performed by carrying out serial dilutions of standard lactate solutions and equine plasma and then analyzing the results with linear regression or paired t test methods.
- A set of 275 blood samples were assessed, collected from horses exercising at different intensity levels. The results were compared using a statistical method known as Bland-Altman analysis.
Results of the study
- The YSI analyzer displayed good linearity for measurement of both lactate standards and equine plasma up to levels of 30 mmol/l, suggesting it is highly accurate. In contrast, while the TDx analyzer also showed good linearity up to 12 mmol/l, it demonstrated decreased accuracy for higher concentrations.
- In terms of parallelism (a method used to ascertain whether the relationship between two different measurements remains constant over a range of different concentrations), the YSI analyzer again performed well compared to the TDx which performed poorly.
- Recovery rates were also compared, and here too, the YSI tool had better performance (101.7 ± 3.4%) than the TDx system (110.6 ± 8.4%). This shows that the YSI recovery rate is within acceptable ranges while TDx’s rate is not.
- The two methods also differ significantly in their within-run and between-run coefficients of variation (CV), which measure method reliability. The YSI system displayed far less variation compared to the TDx
- In a comparison of 275 plasma samples, it was found that the TDx system overestimated lactate readings by a mean value of 1 mmol/l compared to the YSI, adding further to evidence of its inferior performance.
Conclusion
- Based on the extensive testing carried out, the researchers concluded that the YSI method was a reliable and well-performing method for measuring lactate in horse plasma in the range of 0-30 mmol/l.
- The TDx method demonstrated higher variability and generally poorer performance, making it less suitable for use with equine plasma.
Cite This Article
APA
Butudom P, Foreman JH, Kline KH, Whittem EL.
(2011).
Validation and comparison of two methods of measuring lactate in equine plasma.
Equine Vet J Suppl(38), 155-160.
https://doi.org/10.1111/j.2042-3306.2010.00219.x Publication
Researcher Affiliations
- Department of Medicine, Faculty of Veterinary Medicine, Khon Kaen University, Khon Kaen 40002, Thailand. navichaya@kku.ac.th
MeSH Terms
- Animals
- Blood Chemical Analysis / instrumentation
- Blood Chemical Analysis / veterinary
- Horses / blood
- Lactic Acid / blood
- Reproducibility of Results
- Sensitivity and Specificity
Citations
This article has been cited 0 times.Use Nutrition Calculator
Check if your horse's diet meets their nutrition requirements with our easy-to-use tool Check your horse's diet with our easy-to-use tool
Talk to a Nutritionist
Discuss your horse's feeding plan with our experts over a free phone consultation Discuss your horse's diet over a phone consultation
Submit Diet Evaluation
Get a customized feeding plan for your horse formulated by our equine nutritionists Get a custom feeding plan formulated by our nutritionists