Operational Radiation Safety Considerations During Equine Superficial Radiation Treatment Using Portable X-Ray Devices.
Health physics    May 26, 2024   Volume 127, Issue 5 625-632 doi: 10.1097/HP.0000000000001841
Singh A, Johnson T, Wotman K, Leary D.Veterinary personnel exposure due to scattered x rays, when present during superficial radiation treatment (SRT), was evaluated. Veterinary personnel presence is needed during SRT to minimize the degree of required aesthesia to horses. Animal treatment was simulated, and exposure rates determined for 50, 70, and 100 kV x rays using eight different SRT applicators. Exposure rates at the surface of a solid water phantom (SWP) phantom ranged from 3.9 mR h-1 to 396 mR h-1 for 50 kV, from 41 mR h-1 to 2,880 mR h-1 for 70 kV, and from 235 mR h-1 to 7,500 mR h-1, for 100 kV, respectively. A horse pha...