Journal of veterinary internal medicine1995; 9(5); 353-356; doi: 10.1111/j.1939-1676.1995.tb01097.x

A portable blood gas analyzer for equine venous blood.

Abstract: Evaluation of a portable blood gas analyzer, (StatPal II, Unifet, Inc, La Jolla, CA) was performed using tonometered solutions and equine blood. Samples were analyzed by the StatPal II and either an Instrument Laboratory IL1306 (Lexington, MA) or a Radiometer ABL50 blood gas analyzer (Radiometer America Inc., Westlake, OH). Comparison of the StatPal II and the IL1306 was done by analysis of 3 tonometered solutions (acidic, normal, and alkalotic) and 27 equine venous blood samples. Blood pH, PCO2, PO2, and [HCO3] values were altered by IV infusion of 5% sodium bicarbonate or exercising the horses on a treadmill. Comparison of the StatPal II and the Radiometer was performed by analysis of 78 blood samples collected from Standardbred horses before a race. Data were analyzed for the venous blood samples using a paired two-tailed Student's t test and Bland-Altman plots, with significance set at P < .05. The coefficients of variation for pH, Pco2, Po2, and [HCO3-] values of the tonometered solutions analyzed by the StatPal II ranged from 0.067% to 0.087%, 2% to 3.21%, 1.21% to 2.67%, and 0.267% to 0.828%, respectively. Comparison of the equine blood samples analyzed by the StatPal II and the IL1306 demonstrated statistically significant, but clinically irrelevant differences in pH, Pco2, and Po2, but not [HCO3-]. There were statistically significant, but clinically irrelevant differences between the StatPal II and the Radiometer for pH, Pco2, and [HCO3-], but not for Po2. It is concluded that the StatPal II provides reproducible and acceptable analysis of equine venous blood gas samples.
Publication Date: 1995-09-01 PubMed ID: 8531183DOI: 10.1111/j.1939-1676.1995.tb01097.xGoogle Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Comparative Study
  • Journal Article

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

The research study focuses on the evaluation of a portable blood gas analyzer, called StatPal II, using tonometered solutions and equine blood. The researchers carried out multiple comparisons with other gas analyzers to measure its effectiveness. The StatPal II showed variation but was still found to be acceptable in its analysis of equine venous blood gas samples.

Evaluation and Comparison Procedure

  • The evaluation process involved analyzing tonometered solutions, which are specific mixtures, to test StatPal II’s precision and reliability.
  • The solutions used for this process were of three types namely, acidic, normal, and alkalotic to cover the full range of possibilities.
  • After the solutions, 27 equine venous blood samples were also analyzed.
  • To alter the blood pH, PCO2, PO2, and [HCO3] values, two methods were adopted – one by the infusion of 5% sodium bicarbonate and the other via exercising the horses on a treadmill.
  • Aside from the internal testing method, the StatPal II was compared against the Instrument Laboratory IL1306 and a Radiometer ABL50 blood gas analyzer.

Data Analysis and Results

  • Analysis of the data for the venous blood samples was conducted using a paired two-tailed Student’s t-test and Bland-Altman plots. The threshold for significance was set at P < .05.
  • The tonometered solutions displayed coefficients of variation ranging from 0.067% to 0.087% for pH, 2% to 3.21% for Pco2, 1.21% to 2.67% for Po2, and 0.267% to 0.828% for [HCO3-] values when analyzed by the StatPal II.
  • The comparison of the equine blood samples analyzed by the StatPal II and IL1306 recorder statistically significant but clinically irrelevant differences in pH, Pco2, and Po2. However, there were no differences noted for [HCO3-].
  • Similarly, noticeable but clinically unimportant differences were found between the StatPal II and Radiometer for pH, Pco2, and [HCO3-], but not for Po2.

Conclusions

  • The results of these tests led to the conclusion that the StatPal II, despite some minor variations, provides reproducible and acceptable analysis of equine venous blood gas samples.

Cite This Article

APA
Mitten LA, Hinchcliff KW, Sams R. (1995). A portable blood gas analyzer for equine venous blood. J Vet Intern Med, 9(5), 353-356. https://doi.org/10.1111/j.1939-1676.1995.tb01097.x

Publication

ISSN: 0891-6640
NlmUniqueID: 8708660
Country: United States
Language: English
Volume: 9
Issue: 5
Pages: 353-356

Researcher Affiliations

Mitten, L A
  • Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Ohio State University, Columbus 43210, USA.
Hinchcliff, K W
    Sams, R

      MeSH Terms

      • Animals
      • Bicarbonates / blood
      • Blood Gas Analysis / instrumentation
      • Blood Gas Analysis / veterinary
      • Carbon Dioxide / blood
      • Exercise Test
      • Horses / blood
      • Hydrogen-Ion Concentration
      • Oxygen / blood
      • Partial Pressure
      • Reproducibility of Results

      Citations

      This article has been cited 0 times.