Veterinary anaesthesia and analgesia2026; 53(5); 101284; doi: 10.1016/j.vaa.2026.101284

Agreement of a veterinary-specific oscillometric noninvasive arterial blood pressure monitor (SunTech Vet30) with invasive arterial blood pressure in anaesthetised horses.

Abstract: To evaluate agreement between a veterinary-specific oscillometric noninvasive arterial blood pressure monitor and invasive arterial blood pressure (IBP) measurements in anaesthetised horses. Methods: Prospective clinical method-comparison study. Methods: A group of 46 horses undergoing general anaesthesia in dorsal recumbency. Methods: Paired invasive and oscillometric blood pressure measurements were recorded at 15-minute intervals. IBP was measured via a facial artery catheter, and systolic (SAP), diastolic (DAP), and mean arterial pressures (MAP) were recorded. Oscillometric measurements were obtained using a SunTech Vet30 monitor with the cuff positioned at the tail base or distal thoracic limb, according to clinical access, and hydrostatically corrected relative to the IBP transducer level. Agreement was assessed using repeated-measures Bland-Altman analysis, modelling paired differences as oscillometric minus invasive pressure with horse as a random intercept. Results: A total of 311 paired measurements were analysed. Bias was 3.1 mmHg for SAP, -0.4 mmHg for DAP and -4.5 mmHg for MAP. Limits of agreement were -24.6 to 30.8 mmHg for SAP, -24.5 to 23.7 mmHg for DAP and -30.9 to 21.9 mmHg for MAP. Proportional bias was present for DAP and MAP, indicating pressure-dependent differences. Using an IBP MAP of <70 mmHg as the reference definition of hypotension, an oscillometric MAP of <70 mmHg had a sensitivity of 0.774 and a specificity of 0.634. The area under the curve was 0.753 (95% confidence interval 0.632-0.857). Conclusions: The studied oscillometric monitor showed wide limits of agreement with IBP and should not be considered interchangeable with IBP in anaesthetised horses. Conclusions: Although it may provide adjunctive physiological information when IBP monitoring is not feasible, caution is warranted when using oscillometric values to detect hypotension or guide clinical decision making.
Publication Date: 2026-07-08 PubMed ID: 42531665DOI: 10.1016/j.vaa.2026.101284Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Journal Article

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

Objective Overview

  • This study assessed how well a veterinary-specific oscillometric blood pressure monitor (SunTech Vet30) agrees with the gold standard invasive arterial blood pressure (IBP) method in anesthetized horses.

Introduction and Purpose

  • Monitoring arterial blood pressure (ABP) during anesthesia in horses is critical for detecting hypotension and maintaining adequate organ perfusion.
  • Invasive arterial blood pressure (IBP) measurement via arterial catheter is the gold standard but can be technically challenging or not always feasible in clinical settings.
  • Oscillometric noninvasive blood pressure monitors offer an easier alternative but their accuracy and agreement with IBP in horses, especially under anesthesia, needed evaluation.
  • This study aimed to compare readings from the SunTech Vet30 oscillometric monitor, designed for veterinary use, against invasive measurements in anesthetized horses to determine their agreement and reliability.

Methodology

  • A prospective clinical method-comparison study was conducted involving 46 horses undergoing general anesthesia in dorsal recumbency.
  • Paired invasive and oscillometric blood pressure measurements were recorded every 15 minutes throughout anesthesia.
  • Invasive arterial pressure (systolic – SAP, diastolic – DAP, mean – MAP) was measured via a catheter placed in the facial artery, considered a reliable site for IBP in horses.
  • The oscillometric blood pressure was obtained with the SunTech Vet30 monitor using cuffs positioned either at the tail base or on the distal thoracic limb, depending on clinical access.
  • All oscillometric measurements were hydrostatically corrected relative to the level of the IBP transducer to reduce positional bias.
  • Agreement between methods was statistically analyzed using Bland-Altman plots for repeated measures, with differences modeled as oscillometric values minus invasive values, accounting for individual horse variability.

Results

  • A total of 311 paired blood pressure measurements were analyzed.
  • Bias (mean difference) between oscillometric and invasive methods was found to be:
    • SAP: +3.1 mmHg (oscillometric values slightly higher)
    • DAP: -0.4 mmHg (very close)
    • MAP: -4.5 mmHg (oscillometric values slightly lower)
  • The limits of agreement, which describe the range within which 95% of differences fall, were wide:
    • SAP: -24.6 to +30.8 mmHg
    • DAP: -24.5 to +23.7 mmHg
    • MAP: -30.9 to +21.9 mmHg
  • There was proportional bias observed for DAP and MAP, meaning that the differences between the two measurement methods depended on the actual blood pressure level.
  • For clinical detection of hypotension (defined as IBP MAP < 70 mmHg), an oscillometric MAP cutoff of < 70 mmHg produced:
    • Sensitivity: 77.4% (ability to correctly identify hypotensive cases)
    • Specificity: 63.4% (ability to correctly identify non-hypotensive cases)
    • The area under the curve (AUC) for this test was 0.753, indicating moderate overall accuracy.

Conclusions and Clinical Implications

  • The SunTech Vet30 oscillometric monitor showed a substantial amount of variability compared to invasive measurements, with wide limits of agreement.
  • This indicates it should not be used as a direct substitute for invasive blood pressure measurement in anesthetized horses.
  • Although oscillometric measurements can provide helpful supplementary physiological information when IBP monitoring isn’t feasible, they must be interpreted with caution.
  • Care should be taken when using oscillometric blood pressure readings for critical clinical decisions, such as identifying hypotension or adjusting therapy during anesthesia, due to variable accuracy.
  • Overall, invasive blood pressure monitoring remains the preferred method for precise and reliable blood pressure assessment in anesthetized horses whenever possible.

Cite This Article

APA
Hernández-Magaña EZ, Gutiérrez Bautista ÁJ, Viscasillas J, Redondo JI. (2026). Agreement of a veterinary-specific oscillometric noninvasive arterial blood pressure monitor (SunTech Vet30) with invasive arterial blood pressure in anaesthetised horses. Vet Anaesth Analg, 53(5), 101284. https://doi.org/10.1016/j.vaa.2026.101284

Publication

ISSN: 1467-2995
NlmUniqueID: 100956422
Country: United States
Language: English
Volume: 53
Issue: 5
Pages: 101284
PII: S1467-2987(26)00098-X

Researcher Affiliations

Hernández-Magaña, Eva Zoe
  • Departamento de Medicina y Cirugía Animal, Facultad de Veterinaria, Universidad Cardenal, Valencia, Spain. Electronic address: eva.hernandezmagana@uchceu.es.
Gutiérrez Bautista, Álvaro-Jesús
  • Small Animal Clinic, University of Veterinary Medicine Hannover, Foundation, Hannover, Germany.
Viscasillas, Jaime
  • AniCura Valencia Sur Hospital Veterinario, Silla, Valencia, Spain.
Redondo, José I
  • Departamento de Medicina y Cirugía Animal, Facultad de Veterinaria, Universidad Cardenal, Valencia, Spain.

Conflict of Interest Statement

Conflict of interest statement The authors declare that they have no financial or personal conflicts of interest related to this study. The manufacturer of the device had no role in the study design, data collection, analysis, interpretation of the results, or preparation of the manuscript.

Citations

This article has been cited 0 times.