Plasma potassium measurement with a new reagent carrier (Reflotron): comparison with ion-selective electrode results.
Abstract: Potassium concentrations were measured in the plasma of 336 animals with a new reagent carrier (Reflotron; Boehringer Mannheim) K+ and with an ion-specific electrode system: results were highly correlated (r = 0.991; y = 0.993 x + 0.02) and day-to-day coefficient of variation of the new reagent measurements was lower than 2.5 per cent. This system offers a good alternative to the ion-selective electrode system for plasma potassium measurement in veterinary practice.
Publication Date: 1993-01-01 PubMed ID: 8434141DOI: 10.1016/0034-5288(93)90022-8Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
- Comparative Study
- Journal Article
- Analytical Methods
- Animal Health
- Animal Studies
- Biochemistry
- Clinical Findings
- Clinical Pathology
- Clinical Study
- Comparative Study
- Diagnosis
- Diagnostic Technique
- Disease Diagnosis
- Equine Health
- Horses
- Laboratory Methods
- Plasma
- Potassium
- Veterinary Care
- Veterinary Medicine
- Veterinary Practice
- Veterinary Procedure
- Veterinary Research
Summary
This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.
This research article presents the comparative analysis of potassium measurements conducted using a new reagent carrier (Reflotron) and an ion-selective electrode system on animal plasma samples. The research results indicate a high correlation between the two methods and suggest that the Reflotron system could be a reliable alternative for measuring plasma potassium in veterinary practice.
Research Methodology and Data Collection
In this study:
- The researchers measured potassium concentrations in the plasma of 336 animals.
- These measurements were conducted using two methods: a new reagent carrier (Reflotron) and an ion-specific electrode system.
- The Reflotron system used in the study is a product of Boehringer Mannheim, a global pharmaceutical company.
Comparative Analysis Results
The results of this comparative analysis included:
- A high degree of correlation (r = 0.991; y = 0.993 x + 0.02) was observed between the measurements obtained from the two methods.
- In statistical terms, an ‘r’ value close to 1 indicates a strong positive correlation. This suggests that as potassium measurements increase in one method, they similarly increase in the other method.
- The ‘y’ value refers to the regression equation, indicating the relationship between the two sets of measurements.
Variation in Measurements
The research also identified:
- The day-to-day coefficient of variation in measurements obtained with the new reagent was less than 2.5 per cent.
- This low variation is indicative of reliable and consistent results.
Conclusion
Based on the study results:
- The researchers concluded that the Reflotron system offers a good alternative to the ion-selective electrode system for plasma potassium measurement in veterinary practice.
- This conclusion is significant as it indicates potential for adoption of this new method in real-world veterinary diagnostic scenarios. An alternative method that displays such high correlation and low variability is helpful for expanding the options available for such typical tests.
Cite This Article
APA
Braun JP, Carstensen CA.
(1993).
Plasma potassium measurement with a new reagent carrier (Reflotron): comparison with ion-selective electrode results.
Res Vet Sci, 54(1), 127-129.
https://doi.org/10.1016/0034-5288(93)90022-8 Publication
Researcher Affiliations
- Département de Physiopathologie, Ecole Nationale Vétérinaire, Toulouse, France.
MeSH Terms
- Animals
- Blood Chemical Analysis / instrumentation
- Blood Chemical Analysis / veterinary
- Cats
- Cattle
- Dogs
- Electrodes / veterinary
- Horses
- Indicators and Reagents
- Potassium / blood
- Sheep
Citations
This article has been cited 0 times.Use Nutrition Calculator
Check if your horse's diet meets their nutrition requirements with our easy-to-use tool Check your horse's diet with our easy-to-use tool
Talk to a Nutritionist
Discuss your horse's feeding plan with our experts over a free phone consultation Discuss your horse's diet over a phone consultation
Submit Diet Evaluation
Get a customized feeding plan for your horse formulated by our equine nutritionists Get a custom feeding plan formulated by our nutritionists