Veterinary surgery : VS2026; 55(7); 1212-1222; doi: 10.1111/vsu.70126

Radiation exposure to equine surgeons during simulated digital radiography and fluoroscopy-guided fracture fixation of the distal limb.

Abstract: Despite known occupational radiation risks, data on exposure during digital radiography (DR) or fluoroscopy-guided equine fracture repair are lacking. This study quantified radiation doses to a model surgeon's hand, lenses, thyroid, and gonads, and assessed protective effects of radiation-reducing surgical gloves and attenuating ointment. Methods: Comparative, controlled, simulation-based study. Unassigned: Anthropomorphic model. Methods: Distal limb fracture fixation was simulated using a model with dosimeters at the region of the lenses, thyroid, gonads, drill-holding hand, and under a lead apron (fetal dose). Radiation was measured over 10 DR or 10 single-image fluoroscopy exposures, then repeated using radiation-reducing surgical gloves or ointment. Results: All radiation doses were below occupational limits. Dosimeters placed at the level of the gonads (outside the gown) received the highest mean dose (DR: 6.87 ± 0.40 μSv; fluoroscopy: 0.98 ± 0.28 μSv), followed by thyroid (outside the gown) (DR: 2.60 ± 0.52 μSv; fluoroscopy: 0.48 ± 0.08 μSv), hand (DR: 2.02 ± 0.54 μSv; fluoroscopy: 0.32 ± 0.06 μSv) and lenses (DR: 1.25 ± 0.28 μSv; fluoroscopy: 0.23 ± 0.07 μSv). The lowest exposure was measured underneath the gown (fetal dose) (DR: 0.01 ± 0.01 μSv; fluoroscopy: 0.01 ± 0.01 μSv). DR exposure exceeded fluoroscopy at all sites (p < .001). Radiation-reducing surgical gloves significantly lowered hand dose (p < .001), whereas ointment had no effect (p = .54). Off-beam hand exposure from 10 images was approximately one-millionth of the annual dose limit. Conclusions: Image-guided equine distal limb fracture fixation results in minimal hand radiation exposure, relative to other regions of the body, with DR producing higher doses than single-image fluoroscopy. Radiation-reducing surgical gloves significantly lower hand exposure during DR procedures. Conclusions: Radiation-reducing surgical gloves offer additional protection and are recommended during image-guided surgery. Whenever feasible, pulsed fluoroscopy should be preferred over DR.
Publication Date: 2026-06-26 PubMed ID: 42359489DOI: 10.1111/vsu.70126Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Journal Article

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

Radiation exposure to equine surgeons during image-guided fracture fixation of the distal limb is low and well below occupational safety limits, with digital radiography producing higher doses than single-image fluoroscopy. Using radiation-reducing surgical gloves significantly decreases hand exposure during these procedures.

Study Purpose and Background

  • The research aimed to quantify radiation exposure to equine surgeons performing distal limb fracture fixation using digital radiography (DR) or single-image fluoroscopy guidance.
  • Despite acknowledged occupational radiation risks in veterinary surgery, data specifically on exposure during these imaging techniques in equine fracture repair were previously unavailable.
  • The study also assessed the protective benefits of radiation-attenuating surgical gloves and an ointment designed to reduce radiation dose.

Methods

  • A controlled, simulation-based design was used with an anthropomorphic model mimicking the surgeon’s body.
  • Dosimeters were placed at key anatomical sites of radiation concern: lenses (eyes), thyroid gland, gonads, the hand holding the drill, and beneath a lead apron representing fetal exposure.
  • Two types of imaging procedures were tested: 10 exposures of digital radiography and 10 exposures of single-image fluoroscopy.
  • The experiment was repeated twice more, once with radiation-reducing surgical gloves and once using a radiation-attenuating ointment on the hand.

Key Results

  • All measured radiation doses were substantially below the occupational exposure limits broadly accepted for healthcare workers.
  • The highest radiation dose was recorded at the gonads (outside lead protection), followed by the thyroid, hand, and lenses, indicating these areas received varying but relatively low levels of radiation.
  • Radiation measurements (mean ± SD in microsieverts, µSv) for digital radiography versus fluoroscopy were as follows:
    • Gonads: 6.87 ± 0.40 µSv (DR) vs. 0.98 ± 0.28 µSv (fluoroscopy)
    • Thyroid: 2.60 ± 0.52 µSv (DR) vs. 0.48 ± 0.08 µSv (fluoroscopy)
    • Hand: 2.02 ± 0.54 µSv (DR) vs. 0.32 ± 0.06 µSv (fluoroscopy)
    • Lenses: 1.25 ± 0.28 µSv (DR) vs. 0.23 ± 0.07 µSv (fluoroscopy)
    • Fetal dose under lead gown: 0.01 ± 0.01 µSv for both modalities
  • Digital radiography consistently exposed all sites to higher radiation doses than single-image fluoroscopy, with statistically significant differences (p < 0.001).
  • Using radiation-reducing surgical gloves significantly reduced hand dose during digital radiography exposures (p < 0.001), whereas applying the attenuating ointment showed no significant reduction (p = 0.54).
  • Notably, hand exposure off the imaging beam from 10 exposures was about one-millionth of the annual dose limit for radiation workers, indicating very low risk.

Conclusions and Recommendations

  • Radiation exposure to equine surgeons during image-guided distal limb fracture fixation is minimal, particularly to the hands compared with other body regions.
  • Digital radiography produces higher radiation doses than single-image fluoroscopy at all measured locations.
  • Wearing radiation-reducing surgical gloves provides meaningful protection for the surgeon’s hands and is recommended during DR-guided procedures.
  • Whenever possible, pulsed fluoroscopy should be preferred over digital radiography to minimize radiation exposure.
  • Standard radiation protection measures such as lead aprons effectively reduce exposure to internal organs and fetal dose remains negligible.

Implications

  • The study provides important occupational safety data specifically for equine surgeons, supporting informed decisions regarding radiation protection in veterinary orthopedic surgery.
  • Adoption of radiation-reducing gloves and optimization of imaging modalities can further lower surgeons’ radiation risks without compromising procedural success.
  • These findings encourage routine monitoring of radiation exposure and reinforcement of protective practices in equine surgical settings.

Cite This Article

APA
Gaida JL, Stieger-Vanegas SM, Lischer CJ, Duesterdieck-Zellmer KF, Huber MJ, Kamm JL. (2026). Radiation exposure to equine surgeons during simulated digital radiography and fluoroscopy-guided fracture fixation of the distal limb. Vet Surg, 55(7), 1212-1222. https://doi.org/10.1111/vsu.70126

Publication

ISSN: 1532-950X
NlmUniqueID: 8113214
Country: United States
Language: English
Volume: 55
Issue: 7
Pages: 1212-1222

Researcher Affiliations

Gaida, Julia L
  • Department of Clinical Sciences, Carlson College of Veterinary Medicine, Oregon State University, Corvallis, Oregon, USA.
  • Equine Clinic, Faculty of Veterinary Medicine, Freie Universität Berlin, Berlin, Germany.
Stieger-Vanegas, Susanne M
  • Department of Clinical Sciences, Carlson College of Veterinary Medicine, Oregon State University, Corvallis, Oregon, USA.
Lischer, Christoph J
  • Equine Clinic, Faculty of Veterinary Medicine, Freie Universität Berlin, Berlin, Germany.
Duesterdieck-Zellmer, Katja F
  • Department of Clinical Sciences, Carlson College of Veterinary Medicine, Oregon State University, Corvallis, Oregon, USA.
Huber, Michael J
  • Department of Clinical Sciences, Carlson College of Veterinary Medicine, Oregon State University, Corvallis, Oregon, USA.
Kamm, J Lacy
  • Department of Clinical Sciences, Carlson College of Veterinary Medicine, Oregon State University, Corvallis, Oregon, USA.

MeSH Terms

  • Animals
  • Fluoroscopy / veterinary
  • Horses / injuries
  • Horses / surgery
  • Radiation Exposure
  • Occupational Exposure
  • Fractures, Bone / veterinary
  • Fractures, Bone / surgery
  • Fractures, Bone / diagnostic imaging
  • Radiation Dosage
  • Thyroid Gland / radiation effects
  • Fracture Fixation / veterinary
  • Fracture Fixation / methods
  • Radiographic Image Enhancement
  • Forelimb
  • Surgeons

Grant Funding

  • 6748 / American College of Veterinary Surgeons Foundation
  • Carlson College of Veterinary Medicine, Oregon State University

References

This article includes 50 references
  1. Rajaraman P, Doody MM, Yu CL. Journal club: cancer risks in US radiologic technologists working with fluoroscopically guided interventional procedures, 1994‐2008. Am J Roentgenol 2016;206(5):1101‐1109.
  2. Athwal GS, Bueno RA Jr, Wolfe SW. Radiation exposure in hand surgery: mini versus standard C‐arm. J Hand Surg Am 2005;30(6):1310‐1316.
  3. Desouky O, Ding N, Zhou G. Targeted and non‐targeted effects of ionizing radiation. J. Radiat. Res. Appl. Sci. 2015;8(2):247‐254.
  4. Muirhead C, O'hagan J, Haylock R. Mortality and cancer incidence following occupational radiation exposure: third analysis of the National Registry for radiation workers. Br J Cancer 2009;100(1):206‐212.
  5. Richardson DB, Cardis E, Daniels RD. Risk of cancer from occupational exposure to ionising radiation: retrospective cohort study of workers in France, the United Kingdom, and the United States (INWORKS). BMJ 2015;351:h5359.
  6. Commision USNR. Standards for Protection against Radiation Code of Federal Regulations 1991;Title 10(Part 20, Subpart C ‐ Occupational Dose Limits). .
  7. Valentin J, Boice J Jr, Clarke R. Published on behalf of the International Commission on Radiological Protection. Elsevier; 2007.
  8. Kaplan DJ, Patel JN, Liporace FA, Yoon RS. Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle. Patient Saf Surg 2016;10(1):1‐7.
  9. Hafez M, Smith R, Matthews S, Kalap G, Sherman K. Radiation exposure to the hands of orthopaedic surgeons: are we underestimating the risk?. Arch Orthop Trauma Surg 2005;125:330‐335.
  10. Hoffler CE, Ilyas AM. Fluoroscopic radiation exposure: are we protecting ourselves adequately?. J Bone Joint Surg 2015;97(9):721‐725.
  11. Della Vecchia E, Modenese A, Loney T. Risk of cataract in health care workers exposed to ionizing radiation: a systematic review. Med Lav 2020;111(4):269‐284.
  12. Services OSRP. Oregon Administrative Rules, Chapter 333, Division 120, Rule 333–120–0100. 2021.
  13. Chou LB, Lerner LB, Harris AH, Brandon AJ, Girod S, Butler LM. Cancer prevalence among a cross‐sectional survey of female orthopedic, urology, and plastic surgeons in the United States. Womens Health Issues 2015;25(5):476‐481.
  14. Stahl CM, Meisinger QC, Andre MP, Kinney TB, Newton IG. Radiation risk to the fluoroscopy operator and staff. Am J Roentgenol 2016;207(4):737‐744.
  15. Interagency Working Group on Medical Radiation, U.S.. The Use and Care of Personal Protective Equipment for X‐ray Exposure. Environmental Protection Agency; 2015:4.
  16. Naiman JH, Zellner EM, Petrovsky BL. Radiation exposure associated with percutaneous fluoroscopically guided lag screw fixation for sacroiliac luxation in dogs. Vet Surg 2021;50(5):1065‐1075.
  17. Cantlon MB, Ilyas AM. Assessment of radiation protection in hand‐shielding products with mini c‐arm fluoroscopy. Hand 2021;16(4):505‐510.
  18. Ramzan PH, Palmer L. Musculoskeletal injuries in thoroughbred racehorses: a study of three large training yards in Newmarket, UK (2005–2007). Vet J 2011;187(3):325‐329.
  19. Donati B, Fürst A, Hässig M, Jackson MA. Epidemiology of fractures: the role of kick injuries in equine fractures. Equine Vet J 2018;50(5):580‐586.
  20. Steinberg MK. Ermittlung der Strahlenexposition des veterinärmedizinischen Personals bei routinemäßigen Röntgenuntersuchungen in der Pferdepraxis mit Hilfe elektronischer Personendosimeter. .
  21. Dakin S, Lam R, Rees E, Mumby C, West C, Weller R. Technical set‐up and radiation exposure for standing computed tomography of the equine head. Equine Vet Educ 2014;26(4):208‐215.
  22. Thomas HL, Trout DR, Dobson H, McFadden RC. Radiation exposure to personnel during examination of limbs of horses with a portable hand‐held fluoroscopic unit. J Am Vet Med Assoc 1999;215(3):372‐379.
  23. Belotta AF, Mayer MN, Waldner CL. X‐ray tube operators can be exposed to equal or higher scattered radiation doses to the hand as cassette holders during diagnostic radiographic procedures of the equine vertebral column and limbs. Am J Vet Res 2022;83(5):412‐418.
  24. Veitch KE, Szczykutowicz TP, Brounts SH, Ergun DL, Muir P, Loeber SJ. Radiation exposure during simulated equine head and limb fan beam standing computed tomography appears safe for personnel using lead shielding. J Am Vet Med Assoc 2024;1(aop):1‐8:63‐70.
  25. Kim AN, Chang YJ, Cheon BK, Kim JH. How effective are radiation reducing gloves in C‐arm fluoroscopy‐guided pain interventions?. Korean J Pain 2014;27(2):145‐151.
  26. Gaida JL, Steinberg T, Stieger‐Vanegas SM, Merle R, Lischer CJ. Equine standing multidetector computed tomography of the distal thoracic limb and tarsus has a lower cumulative radiation dose than digital radiography. Vet Radiol Ultrasound 2025;66(4):e70049.
  27. Pinheiro J, Bates D, R Core Team. nlme: linear and nonlinear mixed effects models. R package version 3.1‐169. 2026.
  28. Lenth R. emmeans: Estimated marginal means, aka least‐squares means. R package version 1.11.2‐8. 2025.
  29. Kesavachandran CN, Haamann F, Nienhaus A. Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review. Eur J Med Res 2012;17:1‐10.
  30. Riley SA. Radiation exposure from fluoroscopy during orthopedic surgical procedures. Clin Orthop Relat Res 1989;248:257‐260.
  31. Chapman T, Martin DP, Williamson C, Tinsley B, Wang ML, Ilyas AM. Mini C‐arm fluoroscopy: does its configuration matter for radiation exposure to the surgeon?. Hand 2018;13(5):552‐557.
  32. Nelson EM, Monazzam SM, Kim KD, Seibert JA, Klineberg EO. Intraoperative fluoroscopy, portable X‐ray, and CT: patient and operating room personnel radiation exposure in spinal surgery.. Spine J 2014;14(12):2985‐2991.
  33. Singer G. Occupational radiation exposure to the surgeon.. J Am Acad Orthop Surg 2005;13(1):69‐76.
  34. Jones D, Stoddart J. Radiation use in the orthopaedic theatre: a prospective audit.. Occupational Health and Industrial Medicine 1999;2(40):74.
  35. Müller L, Suffner J, Wenda K, Mohr W, Rommens P. Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing.. Injury 1998;29(6):461‐468.
  36. Acho S, Tsoka‐Gwegweni J, Stofile C, Banda K, Theron T, Struwig V. Influence of radiation dose reduction gloves on exposure parameters, entrance dose rates and eye doses of interventionalists during mobile C‐arm fluoroscopic procedures: a phantom study.. Radiography 2023;29(3):539‐545.
  37. United States Environmental Protection Agency IWGoMR. Radiation protection guidance for diagnostic and interventional X‐ray procedures.. Federal Guidance Report No 14, EPA‐402‐R‐10003. Environmental Protection Agency; 2014.
  38. Giordano BD, Baumhauer JF, Morgan TL, Rechtine GR. Patient and surgeon radiation exposure: comparison of standard and mini‐C‐arm fluoroscopy.. J Bone Joint Surg 2009;91(2):297‐304.
  39. Ellis KL, Morton AJ, Hernandez JA, Winter MD. Radiation exposure to personnel obtaining equine appendicular radiographs using a handheld generator.. J Equine Vet Sci 2019;73:70‐74.
  40. Yamashita K, Higashino K, Hayashi H, Hayashi F, Fukui Y, Sairyo K. Pulsation and collimation during fluoroscopy to decrease radiation: a cadaver study.. JBJS Open Access 2017;2(4):e0039.
  41. Vernier TH, Hinson WD, Verpaalen VD. Radiation exposure to the orthopedic surgeon—a dosimetric comparison of two mini C‐arm fluoroscopy models: a pilot study.. J Am Vet Med Assoc 2024;262(3):1‐6.
  42. Kim JH. Three principles for radiation safety: time, distance, and shielding.. Korean J. Pain 2018;31(3):145‐146.
  43. Preston DL, Cullings H, Suyama A. Solid cancer incidence in atomic bomb survivors exposed in utero or as young children.. J Natl Cancer Inst 2008;100(6):428‐436.
  44. Doll R, Wakeford R. Risk of childhood cancer from fetal irradiation.. Br J Radiol 1997;70(830):130‐139.
  45. Chandra V, Dorsey C, Reed AB, Shaw P, Banghart D, Zhou W. Monitoring of fetal radiation exposure during pregnancy.. J Vasc Surg 2013;58(3):710‐714.
  46. Subramanian S, Waller BR, Winders N. Clinical evaluation of a radio‐protective cream for the hands of the pediatric interventional cardiologist.. Catheter Cardiovasc Interv 2017;89(4):709‐716.
  47. Kamusella P, Scheer F, Lüdtke CW, Wiggermann P, Wissgott C, Andresen R. Interventional angiography: radiation protection for the examiner by using lead‐free gloves.. J Clin Diagn Res 2017;11(7):TC26.
  48. Verheyen K, Price J, Lanyon L, Wood J. Exercise distance and speed affect the risk of fracture in racehorses.. Bone 2006;39(6):1322‐1330.
  49. Verheyen K, Wood J. Descriptive epidemiology of fractures occurring in British thoroughbred racehorses in training.. Equine Vet J 2004;36(2):167‐173.
  50. Tyson R, Smiley DC, Pleasant RS, Daniel GB. Estimated operator exposure for hand holding portable X‐ray units during imaging of the equine distal extremity.. Vet Radiol Ultrasound 2011;52(2):121‐124.

Citations

This article has been cited 0 times.