Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide.

Abstract: Plasma adrenocorticotropic hormone (ACTH) concentrations are used to diagnose equine pituitary pars intermedia dysfunction (PPID). The reference chemiluminescent immunoassay for immunoreactive ACTH (iACTH) cross-reacts with corticotropin-like intermediate peptide (CLIP), which is also abnormally high in PPID. Objective: To compare the performance of a novel point-of-care (POC) immunoassay that concurrently measures ACTH and CLIP against the reference iACTH assay. Methods: Twenty-eight university-owned horses (13 PPID, 15 non-PPID). Methods: Prospective observational study. Resting plasma hormone concentrations in all animals were measured by both methods in all seasons over 1 year, and after thyrotropin releasing hormone (TRH)-stimulation testing once. Bias of the POC-ACTH/CLIP assay was assessed using Spearman correlation analysis and Bland-Altman plots, and precision and linearity of the POC-ACTH/CLIP assay were determined. Results: For resting and TRH-stimulated ACTH concentrations, the POC-ACTH/CLIP assay and the reference method had good agreement overall (r ≥ 0.94; P < .001) and in PPID animals (r = 0.93, P < .001). Bias was greater in autumn (33.6-35.6 ± 89.5-96 pg/mL) compared to winter/spring (1.6-11.1 ± 8.4-29.6 pg/mL) and in samples with ACTH concentrations ≥ 90 pg/mL (87.8 ± 116.4 pg/mL). Intra- and inter-assay coefficients of variation for the POC-ACTH/CLIP assay were 8.47% and 14.1%, respectively, and assay linearity was excellent (R2 = 0.991). Conclusions: These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.
Publication Date: 2026-10-11 PubMed ID: 42858361PubMed Central: PMC13653562DOI: 10.1093/jvimsj/aalag253Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Journal Article
  • Validation Study

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

Overview

  • This study evaluated a new point-of-care immunoassay designed to simultaneously measure two hormones, ACTH and CLIP, in horses to diagnose pituitary pars intermedia dysfunction (PPID), comparing its accuracy and precision to the standard laboratory test.

Background

  • Equine pituitary pars intermedia dysfunction (PPID) is a common endocrine disorder in horses, diagnosed primarily through measuring plasma levels of adrenocorticotropic hormone (ACTH).
  • The standard chemiluminescent immunoassay used for ACTH measurement can cross-react with corticotropin-like intermediate peptide (CLIP), another hormone elevated in PPID, potentially affecting accuracy.
  • A novel point-of-care (POC) assay was developed to simultaneously quantify both ACTH and CLIP in plasma samples, potentially improving diagnostic precision and convenience.

Study Objective

  • To validate and compare the performance of the new POC-ACTH/CLIP immunoassay against the established chemiluminescent immunoassay for ACTH (iACTH) in horses.
  • To assess the accuracy, precision, bias, and linearity of the POC assay under various conditions including seasons and after stimulation testing.

Methods

  • Study Design: Prospective observational study involving 28 university-owned adult horses; 13 diagnosed with PPID and 15 without PPID.
  • Sample Collection: Resting plasma hormone levels were measured multiple times across all four seasons over one year, and once after thyrotropin releasing hormone (TRH) stimulation to provoke ACTH secretion.
  • Testing: Each sample was analyzed by both the POC-ACTH/CLIP assay and the reference chemiluminescent iACTH assay.
  • Statistical Analysis:
    • Used Spearman correlation to evaluate the strength of association between the two assays.
    • Bland-Altman plots assessed agreement and bias (difference in hormone values between tests).
    • Intra-assay and inter-assay coefficients of variation (CV) were calculated to evaluate repeatability and reproducibility of the POC assay.
    • Linearity (ability to produce accurate results across a range of concentrations) was evaluated by regression analysis.

Results

  • Correlation: There was a strong positive correlation between POC-ACTH/CLIP and the reference iACTH assay overall (correlation coefficient r ≥ 0.94; P < .001), and specifically in PPID horses (r=0.93; P < .001).
  • Bias:
    • The POC assay tended to show a greater bias (difference from reference values) during autumn, measuring on average 33.6-35.6 pg/mL higher with wide variability (±89.5-96 pg/mL), compared to smaller biases in winter and spring (1.6-11.1 pg/mL ± 8.4-29.6 pg/mL).
    • At higher hormone concentrations (≥ 90 pg/mL), the bias increased substantially (mean difference 87.8 ± 116.4 pg/mL), indicating less precise agreement at elevated levels.
  • Precision:
    • Intra-assay CV (repeatability within the same run) was 8.47%, indicating acceptable repeatability.
    • Inter-assay CV (reproducibility across separate runs) was 14.1%, considered acceptable for clinical assays.
  • Linearity: The assay demonstrated excellent linearity (R² = 0.991), suggesting accurate measurement across a broad range of hormone concentrations.

Conclusions

  • The novel POC-ACTH/CLIP immunoassay performs well in a field setting and correlates closely with the established chemiluminescent assay for ACTH measurement.
  • It offers the advantage of simultaneously measuring both ACTH and CLIP, which may improve diagnostic insight in PPID since both hormones are elevated.
  • Despite overall good agreement, clinicians should be cautious interpreting results during autumn and at higher hormone levels due to increased bias and variability.
  • The assay’s acceptable precision and strong linearity support its use as a convenient and reliable diagnostic tool for equine PPID monitoring.

Cite This Article

APA
Hart KA, Berghaus LJ, Harmon I. (2026). Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide. J Vet Intern Med, 40(5), aalag253. https://doi.org/10.1093/jvimsj/aalag253

Publication

ISSN: 1939-1676
NlmUniqueID: 8708660
Country: England
Language: English
Volume: 40
Issue: 5
PII: aalag253

Researcher Affiliations

Hart, Kelsey A
  • Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States.
Berghaus, Londa J
  • Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States.
Harmon, Ian
  • Zomedica®, Inc; Plymouth, MN, United States.

MeSH Terms

  • Animals
  • Adrenocorticotropic Hormone / blood
  • Horses / blood
  • Immunoassay / veterinary
  • Immunoassay / methods
  • Point-of-Care Systems
  • Female
  • Male
  • Luminescent Measurements / veterinary
  • Pituitary Diseases / veterinary
  • Pituitary Diseases / diagnosis
  • Pituitary Diseases / blood
  • Horse Diseases / blood
  • Horse Diseases / diagnosis
  • Prospective Studies
  • Pituitary Gland, Intermediate

Grant Funding

  • Zomedica, Inc.
  • University of Georgia Foundation

Conflict of Interest Statement

Kelsey Hart serves as a member of the Equine Endocrinology Advisory Board for Boehringer Ingelheim and receives compensation in excess of $5,000/year for this service. Boehringer Ingelheim and Zomedica, Inc., the manufacturer of the POC-ACTH/CLIP assay assessed herein, have a strategic partnership in support of equine endocrine diagnostic testing. Londa Berghaus has no conflicts of interest. Ian Harmon is an employee of Zomedica, Inc.

References

This article includes 23 references
  1. Ireland JL, McGowan CM. Epidemiology of pituitary pars intermedia dysfunction: a systematic literature review of clinical presentation, disease prevalence and risk factors.. Vet J 2018;235:22–33.
    doi: 10.1016/j.tvjl.2018.03.002pubmed: 29704935google scholar: lookup
  2. McGowan TW, Pinchbeck GP, McGowan CM. Prevalence, risk factors and clinical signs predictive for equine pituitary pars intermedia dysfunction in aged horses.. Equine Vet J 2013;45:74–79.
  3. Durham AE, Potier JF, Huber L. The effect of month and breed on plasma adrenocorticotropic hormone concentrations in equids.. Vet J 2022;286:105857.
    doi: 10.1016/j.tvjl.2022.105857pubmed: 35798232google scholar: lookup
  4. Menzies-Gow NJ, Banse HE, Duff A. BEVA primary care clinical guidelines: diagnosis and management of equine pituitary pars intermedia dysfunction.. Equine Vet J 2024;56:220–242.
    doi: 10.1111/evj.14009pubmed: 37795557google scholar: lookup
  5. Stewart AJ, Ireland JL, Durham AE, McGowan CM. Diagnosis of equine pituitary pars intermedia dysfunction.. Vet J 2023;300-302:106036.
    doi: 10.1016/j.tvjl.2023.106036pubmed: 37805159google scholar: lookup
  6. Hart KA, Bamford NJ, Bertin F-R. Recommendations for the diagnosis and management of pituitary pars intemedia dysfunction (PPID).. Accessed June 1, 2025. 2026.
  7. Knowles EJ, Moreton-Clack MC, Shaw S, Harris PA, Elliott J, Menzies-Gow NJ. Plasma adrenocorticotropic hormone (ACTH) concentrations in ponies measured by two different assays suggests seasonal cross-reactivity or interference.. Equine Vet J 2018;50:672–677.
    doi: 10.1111/evj.12797pubmed: 29247453google scholar: lookup
  8. Durham AE. Investigation of peptide cross reactivity in equine plasma using two adrenocorticotropic hormone immunoassays.. Domest Anim Endocrinol 2025;93:106959.
  9. Carmalt JL, Mortazavi S, McOnie RC, Allen AL, Unniappan S. Profiles of pro-opiomelanocortin and encoded peptides, and their processing enzymes in equine pituitary pars intermedia dysfunction.. PloS One 2018;13:e0190796.
  10. Knowles EJ, Hyde C, Harris PA, Elliott J, Menzies-Gow NJ. Short communication: identification of equine corticotropin-like intermediate lobe peptide (CLIP) binding to an adrenocortipcotrophic hormone (ACTH) assay capture antibody.. Domest Anim Endocrinol 2023;83:106785.
  11. Bamford NJ, Stewart AJ, El-Hage CM, Bertin FR, Bailey SR. Investigation of breed differences in plasma adrenocorticotropic hormone concentrations among healthy horses and ponies.. Vet J 2023;296-297:105995.
    doi: 10.1016/j.tvjl.2023.105995pubmed: 37207985google scholar: lookup
  12. Vaughn SA, Lemons MB, Hart KA. The effect of season and breed on hypothalamic-pituitary-adrenal axis hormones, metabolic hormones, and oxidative markers in ponies and horses.. J Vet Intern Med 2025;39:e70047.
    doi: 10.1111/jvim.70047pmc: PMC11884412pubmed: 40048369google scholar: lookup
  13. Peterson ME, Dougherty E, Rishniw M. Evaluation of a novel, sensitive thyroid-stimulating hormone assay as a diagnostic test for thyroid disease in cats.. Am J Vet Res 2024;85:1-11.
    doi: 10.2460/ajvr.23.12.0278pubmed: 38382201google scholar: lookup
  14. Vaughn SA, Norton NA, Hart KA. Circulating hypothalamic-pituitary-adrenal Axis hormones and insulin concentrations in horses and ponies.. J Equine Vet Sci 2022;111:103810.
    doi: 10.1016/j.jevs.2021.103810pubmed: 35219029google scholar: lookup
  15. Neufang L, Ramos J, Eda S, Flatland B, Giori L. Initial development of a rapid, portable, stall-side ELISA for the measurement of equine adrenocorticotropic hormone.. J Vet Diagn Invest 2025;37:208–211.
    doi: 10.1177/10406387241285453pmc: PMC11559900pubmed: 39320416google scholar: lookup
  16. Steel NL, Ireland JL, McGowan CM. Management of pituitary pars intermedia dysfunction in practice: a clinical audit.. Vet J 2022;289:105899.
    doi: 10.1016/j.tvjl.2022.105899pubmed: 36162625google scholar: lookup
  17. Adams AA, Siard-Altman MH, Reedy SE. Evaluation of seasonal influences on adrenocorticotropic hormone response to the thyrotropin-releasing hormone stimulation test and its accuracy for diagnosis of pituitary pars intermedia dysfunction.. Vet J 2023;300-302:106035.
    doi: 10.1016/j.tvjl.2023.106035pubmed: 37802466google scholar: lookup
  18. Auvinen JRE, Stapley ED, Bertin FR. The ACTH ratio validates the thyrotropin-releasing hormone stimulation test to diagnose pituitary pars intermedia dysfunction but does not enhance performance.. Am J Vet Res 2026;87:1-5.
    doi: 10.2460/ajvr.25.10.0359pubmed: 41468738google scholar: lookup
  19. Durham AE. The effect of pergolide mesylate on adrenocorticotrophic hormone responses to exogenous thyrotropin releasing hormone in horses.. Vet J 2022;285:105831.
    doi: 10.1016/j.tvjl.2022.105831pubmed: 35477010google scholar: lookup
  20. Schott HC 2nd, Strachota JR, Marteniuk JV, Refsal KR. Long-term response of equids with pituitary pars intermedia dysfunction to treatment with pergolide.. J Vet Intern Med 2025;39:e70109.
    doi: 10.1111/jvim.70109pmc: PMC12046498pubmed: 40317948google scholar: lookup
  21. Banse HE, Schultz N, McCue M, Geor R, McFarlane D. Comparison of two methods for measurement of equine adrenocorticotropin.. J Vet Diagn Invest 2018;30:233–237.
    doi: 10.1177/1040638717752216pmc: PMC6505866pubmed: 29284383google scholar: lookup
  22. Pinn-Woodcock TL, Llanos-Soto SG, Ivanek R. Seasonal elevation in equine adrenocorticotropic hormone occurs throughout the contiguous United States and is influenced by latitude and age.. J Am Vet Med Assoc 2025;263:1380–1389.
    doi: 10.2460/javma.25.04.0223pubmed: 40749730google scholar: lookup
  23. Linnet K. Necessary sample size for method comparison studies based on regression analysis.. Clin Chem 1999;45:882–894.
    doi: 10.1093/clinchem/45.6.882pubmed: 10351998google scholar: lookup

Citations

This article has been cited 0 times.