Abstract: Plasma adrenocorticotropic hormone (ACTH) concentrations are used to diagnose equine pituitary pars intermedia dysfunction (PPID). The reference chemiluminescent immunoassay for immunoreactive ACTH (iACTH) cross-reacts with corticotropin-like intermediate peptide (CLIP), which is also abnormally high in PPID. Objective: To compare the performance of a novel point-of-care (POC) immunoassay that concurrently measures ACTH and CLIP against the reference iACTH assay. Methods: Twenty-eight university-owned horses (13 PPID, 15 non-PPID). Methods: Prospective observational study. Resting plasma hormone concentrations in all animals were measured by both methods in all seasons over 1 year, and after thyrotropin releasing hormone (TRH)-stimulation testing once. Bias of the POC-ACTH/CLIP assay was assessed using Spearman correlation analysis and Bland-Altman plots, and precision and linearity of the POC-ACTH/CLIP assay were determined. Results: For resting and TRH-stimulated ACTH concentrations, the POC-ACTH/CLIP assay and the reference method had good agreement overall (r ≥ 0.94; P < .001) and in PPID animals (r = 0.93, P < .001). Bias was greater in autumn (33.6-35.6 ± 89.5-96 pg/mL) compared to winter/spring (1.6-11.1 ± 8.4-29.6 pg/mL) and in samples with ACTH concentrations ≥ 90 pg/mL (87.8 ± 116.4 pg/mL). Intra- and inter-assay coefficients of variation for the POC-ACTH/CLIP assay were 8.47% and 14.1%, respectively, and assay linearity was excellent (R2 = 0.991). Conclusions: These data demonstrate acceptable field performance for the POC-ACTH/CLIP assay and good agreement with the chemiluminescent iACTH immunoassay, with a larger bias in samples with higher hormone concentrations.
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.
Overview
This study evaluated a new point-of-care immunoassay designed to simultaneously measure two hormones, ACTH and CLIP, in horses to diagnose pituitary pars intermedia dysfunction (PPID), comparing its accuracy and precision to the standard laboratory test.
Background
Equine pituitary pars intermedia dysfunction (PPID) is a common endocrine disorder in horses, diagnosed primarily through measuring plasma levels of adrenocorticotropic hormone (ACTH).
The standard chemiluminescent immunoassay used for ACTH measurement can cross-react with corticotropin-like intermediate peptide (CLIP), another hormone elevated in PPID, potentially affecting accuracy.
A novel point-of-care (POC) assay was developed to simultaneously quantify both ACTH and CLIP in plasma samples, potentially improving diagnostic precision and convenience.
Study Objective
To validate and compare the performance of the new POC-ACTH/CLIP immunoassay against the established chemiluminescent immunoassay for ACTH (iACTH) in horses.
To assess the accuracy, precision, bias, and linearity of the POC assay under various conditions including seasons and after stimulation testing.
Methods
Study Design: Prospective observational study involving 28 university-owned adult horses; 13 diagnosed with PPID and 15 without PPID.
Sample Collection: Resting plasma hormone levels were measured multiple times across all four seasons over one year, and once after thyrotropin releasing hormone (TRH) stimulation to provoke ACTH secretion.
Testing: Each sample was analyzed by both the POC-ACTH/CLIP assay and the reference chemiluminescent iACTH assay.
Statistical Analysis:
Used Spearman correlation to evaluate the strength of association between the two assays.
Bland-Altman plots assessed agreement and bias (difference in hormone values between tests).
Intra-assay and inter-assay coefficients of variation (CV) were calculated to evaluate repeatability and reproducibility of the POC assay.
Linearity (ability to produce accurate results across a range of concentrations) was evaluated by regression analysis.
Results
Correlation: There was a strong positive correlation between POC-ACTH/CLIP and the reference iACTH assay overall (correlation coefficient r ≥ 0.94; P < .001), and specifically in PPID horses (r=0.93; P < .001).
Bias:
The POC assay tended to show a greater bias (difference from reference values) during autumn, measuring on average 33.6-35.6 pg/mL higher with wide variability (±89.5-96 pg/mL), compared to smaller biases in winter and spring (1.6-11.1 pg/mL ± 8.4-29.6 pg/mL).
At higher hormone concentrations (≥ 90 pg/mL), the bias increased substantially (mean difference 87.8 ± 116.4 pg/mL), indicating less precise agreement at elevated levels.
Precision:
Intra-assay CV (repeatability within the same run) was 8.47%, indicating acceptable repeatability.
Inter-assay CV (reproducibility across separate runs) was 14.1%, considered acceptable for clinical assays.
Linearity: The assay demonstrated excellent linearity (R² = 0.991), suggesting accurate measurement across a broad range of hormone concentrations.
Conclusions
The novel POC-ACTH/CLIP immunoassay performs well in a field setting and correlates closely with the established chemiluminescent assay for ACTH measurement.
It offers the advantage of simultaneously measuring both ACTH and CLIP, which may improve diagnostic insight in PPID since both hormones are elevated.
Despite overall good agreement, clinicians should be cautious interpreting results during autumn and at higher hormone levels due to increased bias and variability.
The assay’s acceptable precision and strong linearity support its use as a convenient and reliable diagnostic tool for equine PPID monitoring.
Cite This Article
APA
Hart KA, Berghaus LJ, Harmon I.
(2026).
Validation of a point-of-care assay for simultaneous quantification of equine adrenocorticotropic hormone and corticotropin-like intermediate peptide.
J Vet Intern Med, 40(5), aalag253.
https://doi.org/10.1093/jvimsj/aalag253
Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States.
Berghaus, Londa J
Department of Large Animal Medicine, University of Georgia College of Veterinary Medicine, Athens, GA 30602, United States.
Harmon, Ian
Zomedica®, Inc; Plymouth, MN, United States.
MeSH Terms
Animals
Adrenocorticotropic Hormone / blood
Horses / blood
Immunoassay / veterinary
Immunoassay / methods
Point-of-Care Systems
Female
Male
Luminescent Measurements / veterinary
Pituitary Diseases / veterinary
Pituitary Diseases / diagnosis
Pituitary Diseases / blood
Horse Diseases / blood
Horse Diseases / diagnosis
Prospective Studies
Pituitary Gland, Intermediate
Grant Funding
Zomedica, Inc.
University of Georgia Foundation
Conflict of Interest Statement
Kelsey Hart serves as a member of the Equine Endocrinology Advisory Board for Boehringer Ingelheim and receives compensation in excess of $5,000/year for this service. Boehringer Ingelheim and Zomedica, Inc., the manufacturer of the POC-ACTH/CLIP assay assessed herein, have a strategic partnership in support of equine endocrine diagnostic testing. Londa Berghaus has no conflicts of interest. Ian Harmon is an employee of Zomedica, Inc.
Hart KA, Bamford NJ, Bertin F-R. Recommendations for the diagnosis and management of pituitary pars intemedia dysfunction (PPID).. Accessed June 1, 2025. 2026.
Vaughn SA, Lemons MB, Hart KA. The effect of season and breed on hypothalamic-pituitary-adrenal axis hormones, metabolic hormones, and oxidative markers in ponies and horses.. J Vet Intern Med 2025;39:e70047.
Peterson ME, Dougherty E, Rishniw M. Evaluation of a novel, sensitive thyroid-stimulating hormone assay as a diagnostic test for thyroid disease in cats.. Am J Vet Res 2024;85:1-11.