Analyze Diet

Evaluation of the Sidekick point-of-care progesterone lateral-flow assay for use in equine reproductive management.

Abstract: A point-of-care (POC; stall-side) test for progesterone that delivers rapid, accurate results has the potential to transform equine reproductive care. We evaluated the analytical and clinical performance of a POC device (Sidekick) that incorporates a progesterone lateral-flow assay (LFA) with a reader, and compared it with established methods (radioimmunoassay [RIA] and chemiluminescence immunoassay [CLIA]). Analytical performance was assessed using serially diluted mare plasma (0-8 ng/mL progesterone) to estimate precision and percent recovery. Clinical accuracy was assessed by comparison with RIA and CLIA on 19 paired serum samples. Agreement was evaluated by regression analyses and ordinal concordance across clinically relevant progesterone bands. The LFA had excellent recovery (90-102%) across 4-8 ng/mL, strong repeatability (CV 4-15%), and high reproducibility between 2 laboratories (CV 9.9%). Ordinal agreement with reference standards was near-perfect in spiked samples (κ = 0.94) and substantial in clinical samples versus both RIA (κ = 0.77) and CLIA (κ = 0.79). Cross-reactivity with the synthetic progestin, altrenogest, was tested at pharmacologic concentrations; no cross-reactivity was detected. Limitations of our study include an observed bias, potentially caused by the matrix differing between methods (serum vs. plasma). Additionally, a narrow validated range could cause underestimation of high concentrations. The LFA in the Sidekick device provided rapid, accurate, and clinically meaningful progesterone results in mares across 4-8 ng/mL, which would enable on-the-spot decisions and avoid delays from sample processing, transport, and result reporting.
Publication Date: 2026-07-08 PubMed ID: 42417414PubMed Central: PMC13346099DOI: 10.1177/10406387261460834Google Scholar: Lookup
The Equine Research Bank provides access to a large database of publicly available scientific literature. Inclusion in the Research Bank does not imply endorsement of study methods or findings by Mad Barn.
  • Journal Article
  • Evaluation Study

Summary

This research summary has been generated with artificial intelligence and may contain errors and omissions. Refer to the original study to confirm details provided. Submit correction.

Research Overview

  • This study evaluated the effectiveness of the Sidekick, a point-of-care progesterone test device for horses, comparing its accuracy and reliability with established laboratory methods.
  • The goal was to determine if the Sidekick could provide rapid, accurate progesterone measurements to improve equine reproductive management directly at the stall-side.

Purpose and Importance of the Study

  • Progesterone measurement is critical in managing equine reproduction, as it helps in monitoring the mare’s reproductive cycle and pregnancy status.
  • Traditional progesterone assays like radioimmunoassay (RIA) and chemiluminescence immunoassay (CLIA) require laboratory processing, causing delays.
  • A rapid, accurate point-of-care (POC) device could enable immediate clinical decisions, improving reproductive care efficiency.

The Sidekick Device and Progesterone Lateral-Flow Assay (LFA)

  • The Sidekick device uses a progesterone lateral-flow assay, a rapid test format similar to a pregnancy test, combined with an electronic reader for quantifiable results.
  • This setup offers quick progesterone level readouts at the stall-side without needing lab equipment.

Analytical Performance Evaluation

  • Researchers tested serial dilutions of mare plasma with known progesterone concentrations (ranging from 0 to 8 ng/mL) to assess the Sidekick’s accuracy and precision.
  • Key metrics included:
    • Percent recovery: How close the test’s result was to the true concentration (Sidekick showed 90-102% recovery between 4-8 ng/mL, indicating high accuracy).
    • Repeatability: Consistency of results when the same sample is tested multiple times (Coefficient of Variation [CV]: 4-15%, demonstrating strong repeatability).
    • Reproducibility: Consistency of results across two different labs (CV: 9.9%, indicating good reproducibility).

Clinical Accuracy and Agreement with Established Methods

  • Nineteen paired serum samples were tested by the Sidekick device and compared to RIA and CLIA results.
  • Agreement was measured through regression analyses and ordinal concordance (κ, kappa statistic) across progesterone ranges relevant to clinical decisions.
  • Findings demonstrated:
    • Near-perfect agreement with spiked samples (κ = 0.94), indicating the Sidekick accurately detected known progesterone levels.
    • Substantial agreement with clinical samples compared to RIA (κ = 0.77) and CLIA (κ = 0.79), showing the device’s results closely matched established assays in real-world samples.

Cross-Reactivity Testing

  • The study assessed whether the Sidekick test reacted with altrenogest, a synthetic progestin drug commonly administered to mares.
  • No cross-reactivity was detected at pharmacologic concentrations, meaning the Sidekick specifically measured natural progesterone without interference from altrenogest.

Limitations Identified

  • An observed bias was noted that could arise from differences in sample matrices: the Sidekick used plasma, while the reference methods used serum.
  • The narrow validated range of the Sidekick (up to 8 ng/mL) might underestimate progesterone levels when concentrations are higher than this range.
  • These factors should be considered when interpreting results, particularly at very high progesterone levels or comparing plasma versus serum.

Conclusions and Implications for Equine Reproductive Management

  • The Sidekick lateral-flow device reliably provides rapid and accurate progesterone measurements within the clinically important concentration range (4-8 ng/mL).
  • Its rapid results facilitate on-the-spot reproductive decisions by veterinarians or breeders.
  • Using the Sidekick could reduce delays caused by sample processing and laboratory testing, potentially improving reproductive outcomes in horses.

Cite This Article

APA
McDowell SH, McLain A, Moore L, Ledgerwood CJ, Bailie E, Moore T. (2026). Evaluation of the Sidekick point-of-care progesterone lateral-flow assay for use in equine reproductive management. J Vet Diagn Invest, 38(6), 1238-1245. https://doi.org/10.1177/10406387261460834

Publication

ISSN: 1943-4936
NlmUniqueID: 9011490
Country: United States
Language: English
Volume: 38
Issue: 6
Pages: 1238-1245
PII: 10406387261460834

Researcher Affiliations

McDowell, Sara H
  • Future Medicines Institute, Ulster University, Belfast, Northern Ireland, United Kingdom.
McLain, Andrew
  • Future Medicines Institute, Ulster University, Belfast, Northern Ireland, United Kingdom.
Moore, Luca
  • University of York, Heslington, York, United Kingdom.
Ledgerwood, Craig J
  • Future Medicines Institute, Ulster University, Belfast, Northern Ireland, United Kingdom.
Bailie, Elizabeth
  • Future Medicines Institute, Ulster University, Belfast, Northern Ireland, United Kingdom.
Moore, Tara
  • Future Medicines Institute, Ulster University, Belfast, Northern Ireland, United Kingdom.

MeSH Terms

  • Animals
  • Horses / blood
  • Horses / physiology
  • Progesterone / blood
  • Female
  • Point-of-Care Systems
  • Reproducibility of Results
  • Immunoassay / veterinary
  • Immunoassay / methods
  • Rapid Diagnostic Tests
  • Radioimmunoassay / veterinary
  • Pregnancy

Conflict of Interest Statement

Declaration of conflicting interestsThe authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

References

This article includes 20 references
  1. American Association of Equine Practitioners. Diagnostic sample collection, handling, and transport guide. 2024.
  2. Galvin P. The evaluation of three diagnostic tests for the detection of equine influenza nucleoprotein in nasal swabs. Influenza Other Respir Viruses 2014;8:376–383.
    pmc: PMC4181487pubmed: 24512560
  3. Henderson KM, Eayrs K. Pregnancy status determination in mares using a rapid lateral flow test for measuring serum oestrone sulphate. N Z Vet J 2004:52;193–196.
    pubmed: 15726130
  4. Khelifa L. Lateral flow assays for hormone detection. Lab Chip 2022;22:2451–2475.
    pubmed: 35713489
  5. Nagy P. Progesterone determination in equine plasma using different immunoassays. Acta Vet Hung 1998;46:501–513.
    pubmed: 9713151
  6. Orakpoghenor O. Problems in veterinary pathology: a focus on diagnosis. Eur J Sci Res Rev 2024;1:103–111.
  7. Relave F. Accuracy of a rapid enzyme-linked immunosorbent assay to measure progesterone in mares. Can Vet J 2007;48:823–826.
    pmc: PMC1914315pubmed: 17824324
  8. Ross GMS. Unraveling the hook effect: a comprehensive study of high antigen concentration effects in sandwich lateral flow immunoassays. Anal Chem 2020;92:15587–15595.
    pmc: PMC7711776pubmed: 33185097
  9. Safronova VA. Lateral flow immunoassay for progesterone detection. Moscow Univ Chem Bull 2012;67:241–248.
  10. Satué K, Gardón JC. Physiological and clinical aspects of the endocrinology of the estrous cycle and pregnancy in mares. Animal Reproduction in Veterinary Medicine IntechOpen, 2020:90387.
  11. Schwartz D. Analytical validation of a new point-of-care assay for serum amyloid A in horses. Equine Vet J 2018;50:678–683.
    pubmed: 29344980
  12. Seren E. Concentrations of progesterone, 17α-hydroxyprogesterone and 20α-dihydroprogesterone in the plasma of mares during pregnancy and at parturition. J Reprod Fertil 1981;63:443–448.
    pubmed: 7299748
  13. Shikichi M. Abnormal pregnancies associated with deviation in progestin and oestrogen profiles in late pregnant mares: a diagnostic aid. Theriogenology 2017;98:75–81.
    pubmed: 28601159
  14. Squires EL. Use of progestins in open and pregnant mares. Anim Reprod Sci 1993;33:183–193.
  15. Staempfli SA. Do I give altrenogest to prevent pregnancy failure?. Equine Vet Educ 2015;27:498–499.
  16. Tholen DW. Evaluation of precision performance of quantitative measurement methods; approved guideline—second edition. CLSI document EP5-A2. CLSI, 2003.
  17. Viner M. Comparison of serum amyloid A in horses with infectious and non-infectious respiratory diseases. J Equine Vet Sci 2017;49:11–13.
  18. Wagner L. Low progesterone concentration in early pregnancy is detrimental to conceptus development and pregnancy outcome in horses. Anim Reprod Sci 2023;257:107334.
    pubmed: 37725863
  19. World Organisation for Animal Health. Manual of diagnostic tests and vaccines for terrestrial animals. WOAH, 2019.
  20. Wynn MAA. A comparison of progesterone assays for determination of peripheral pregnane concentrations in the late pregnant mare. Theriogenology 2018;106:127–133.
    pubmed: 29059599

Citations

This article has been cited 0 times.